FOXA100
FOXA100 is a microprocessor based
thermal conductivity instrument for testing in accordance with
ASTM C 177 and ISO 8302. Minimum sample size is 65 mm diameter.
The actual measuring area is 36.9 mm. The instrument accepts
samples up to 150 mm square and up to 28 mm thick. If 150 mm
square samples are used the sample acts as part of the guarding
system by covering the rest of the plate. Samples smaller than
150 mm square require surround guards to cover the rest of the
guard zone. The instrument produces conductivity results
accurate to within 1% with 0.2% repeatability and 0.5%
reproducibility. Both instrument plates utilize solid state
heating/cooling and operate between 0°C and 80°C (32°F and
176°F). (Temperatures of 145°C are available in advanced
version.)
The Main Heater Assembly is the heart of the Absolute System. It
consists of three concentric heater systems in one. The central
main heater where the actual conductivity measurement is made is
36.9 mm in diameter and surrounded by two independent concentric
heater rings 12.5 mm wide. The power dissipated in the heater is
measured to arrive at the conductivity. The temperature of the
guard area rings is matched with the temperature of the main
heater to minimize lateral heat flow. An additional guarding
system of heating\cooling modules provides isothermal
temperature over the rest of the 150 mm plate area.
A Heat Flow Meter is used to control temperature of the bottom
surface plate to assure that all power generated in the center
section of the main heater will be driven through the measuring
area of the sample.
FOXA100 utilizes state of the art analog and digital electronics
developed specifically for thermal conductivity testing
instruments and refined over years of use in our FOX Heat Flow
Meters. It is also very simple to operate. Step by step
instructions are displayed on clear four line LCD display.
Plate spacing is automatically adjusted to the desired specimen
thickness (accurate to within +/-0.025mm ( +/-0.001")). This
accuracy is achieved through the use of a sophisticated digital
plate position control system with readout in all four corners
of the specimen.
Input power can be either 115V or 220V 50/60Hz user switchable
at 700VA. Water at 18°C (65°F) or lower is also required.
Instrument dimensions: 330 mm wide by 620 mm deep and 465 mm
high (with Bell Jar).