spektral@spektral.com.tr

|

+90 533 485 47 70

TR

  • ThermCon THS 200 Thermal Conductivity Tester
  • ThermCon THS 200 Thermal Conductivity Tester

ThermCon THS 200 Thermal Conductivity Tester

  • Product Introduction

    The THS 200 is a newly launched thermal conductivity tester by ThermCon Scientific Instrument, which adopts the transient heat source method. It features a fast measurement speed, capable of calculating thermal conductivity within 160 seconds, and can automatically adjust reference and power levels, significantly enhancing experimental efficiency. It also offers a wide measurement range, no restriction on sample morphology, bidirectional operation system, and color touchscreen interface.

  • Testing Principle

    The transient planar heat source method is based on the principle of one-dimensional unsteady thermal conduction on a plane: in an infinite medium, the plane heat source generates a dynamic temperature field in the medium after receiving an instantaneous heating pulse under initial thermal equilibrium. The sample’s thermal conductivity and thermal diffusivity coefficient are calculated by fitting the temperature data generated during heat transfer and fitting a functional curve. During the test, a constant direct current is applied to the heat source, increasing its surface temperature and resistance, leading to an unbalanced bridge test system and generating potential changes. From the changes in electrical parameters, the temperature rise over time is obtained.

  • Testing Scope

    The ThermCon Scientific Thermal Conductivity Tester is used to measure the thermal conductivity of various materials, including bulk solids, metals, liquids, powders, colloids, pastes, films, and insulation materials.

  • Performance Advantages

    1. Integrated model design is lighter and more convenient.
    2. Fully automatic comparison and power adjustment significantly enhance experimental efficiency.
    3. Upgraded instrument software improves user experience and expands specific heat capacity calculations.
    4. Enhanced analysis software provides stronger data analysis; thermal conductivity is calculated directly.
    5. Bidirectional operation system: both software and device can control operations, offering greater convenience.
    6. Wide measurement range accommodates testing needs of different material types.

  • Technical Specifications

    Parameter

    Specification

    Measurement range

    0.00001—50 W/(m·K)

    Temperature range

    -30°C to 200°C

    Test probes

    2 high-precision probes; optional anti-magnetic probes

    Accuracy

    0,001 W /mK 

    Repeatability error

    ≤1%

    Measurement time

    ≤160 seconds

    Total power consumption

    <300 W

    Solid sample size (Probe 1)

    ≥15 mm × 15 mm × 7.5 mm

    Solid sample size (Probe 2)

    ≥30 mm × 30 mm × 15 mm

    Liquid/powder sample capacity

    Matching sample holder, capacity: 18 ml

  • Accessories

    • 1.png
    • 2.png
    • 3.png

  • Technical Specifications

    Temperature range: Room temperature
    Thermal conductivity range: 0.001 50W/m.K
    Test probe: 2 high-precision probes, optional anti magnetization probe
    Precision: 0.001W/ ( m*K )
    Accuracy: ± 5 %
    Reproducibility ± 3 %