  • General Information

    Hot-wall retort furnaces are perfectly suited for various heat treatment processes requiring a defined protective or reaction gas atmosphere with slight overpressure. Depending on the temperature, they are equipped with direct or indirect heating. These compact, gas-tight furnaces can also be designed for heat treatment under vacuum conditions up to 600 °C.

    The furnace chamber consists of a gas-tight retort with water cooling around the door to protect the special sealing. With corresponding safety technology, retort furnaces are also suitable for applications under reaction gases such as hydrogen, or—when combined with the IDB package—for inert debinding or pyrolysis processes.

  • Model Versions

    Models NRA ../06 – Tmax 600 °C

    - Heating elements located inside the retort
    - Retort made of 1.4571 (X6CrNiMoTi 17-12-2)
    - Air circulation fan and baffle for directed gas flow
    - Insulation made of mineral wool
    - Furnace temperature control with measurement inside the retort

    Models NRA ../09 – Tmax 900 °C

    - Based on NRA ../06, with differences:
    - Outside heating with heating elements around the retort
    - Retort made of 1.4828 (X15CrNiSi 20-12)
    - Multi-layer refractory insulation and microporous panel material
    - Furnace temperature control with measurement outside the retort

    Models NR ../11 – Tmax 1100 °C

    - Based on NRA ../09, with differences:
    - Retort made of 1.4841 (X15CrNiSi 25-21)
    - Without gas circulation and baffles
    - Welded support brackets

  • Standard Equipment

    - Compact design with integrated control and gas supply (up to NR(A) 700/..)
    - Exclusive use of insulation materials not classified under EC Regulation No. 1272/2008 (CLP)
    *Note: Alumino-silicate wool (refractory ceramic fiber, RCF), classified as potentially carcinogenic, is not used.*
    - Swivel door hinged on the right side
    - Open cooling water system
    - Multi-zone heating control
    - Temperature uniformity up to ±8 °C according to DIN 17052-1 (in empty work space)
    - Gas supply system for one non-flammable protective or reaction gas, with flow meter and magnetic valve
    - Defined application within constraints of operating instructions
    - NTLog Basic for Nabertherm controller: process data recording with USB flash drive
    - Controller with touch operation (P570)

  • Additional Equipment (Options)

    - Upgrade for other non-flammable gases
    - Mass flow controller (MFC)
    - Process controls H3700, H1700 (PLC) with remote maintenance module
    - Charge control with temperature measurement inside and outside the retort
    - Indirect and/or direct cooling
    - Heat exchanger with closed-loop water circuit for door cooling
    - Oxygen sensor and dew point sensor
    - Parallel swing door or electric bayonet catch
    - Retort made of 2.4633 for Tmax 1150 °C
    - External switchgear with or without cabinet cooling
    - Charge support or custom-built charging trolley
    - Temperature uniformity optimized to ±5 °C (DIN 17052 or AMS 2750G) in empty work space
    - Process control and documentation via VCD software package or Nabertherm Control Center (NCC)